Ultrasonic Specifications
Measurement Modes:
Test Range:
0 - 1mm (0.05in) up to 0 - 20000mm (800in.)
at steel velocity: Variable in 1,2,5 sequence
or continuously in 1mm
(0.05in) increments. Also from 1 to 5000.
Velocity:
256 to 16000m/s continuously variable.
Probe
Zero: 0 to 999.999μs, continuously
variable.
Delay:
Calibrated delay from 0 -20000mm in 0.05mm
steps at steel
velocity (0-400in. in 0.002in. steps).
Gain:
0 to 110dB. Adjustable in 0.5, 1, 2, 6,
10, 14 and 20dB steps. Direct access to
gain control at all times.
Test
Modes: Pulse echo and transmit/receive.
Pulser: 100V - 450V square
wave pulser. Pulse width from Spike to 2000ns
duration - rise/fall times <10ns into 50
ohms at 200V: Width adjustable in 2% of
nominal width, minimum 1ns maximum 40ns.
ActiveEdge™
Unique active pulse control for enhanced
near surface resolution and signal response.
Replaces traditional damping control.
P.R.F: Selectable 5 to
100Hz in 5Hz steps then 100 to 500Hz in
50Hz steps.
Update
Rate: 60Hz (NTSC Mode); 50Hz (PAL
Mode).
Rectification:
Full wave, positive or negative half wave
and unrectified rf.
Frequency
Range: 5 narrow bands centred at
0.25MHz, 0.5MHz, 1MHz, 2 to 2.25MHz, 4 to
5MHz. Broad band at 2MHz to 22MHz (-6dB)
and 1MHz to
35MHz (-20dB).
System
Linearity: Vertical = 1% Full Screen
Height (FSH). Amplifier Accuracy ± 0.1dB.
Horizontal ±0.4% Full Screen Width (FSW).
Reject:
80% linear reject. LED warning light when
selected.
Units:
Metric (mm), inch (in) or time (μs).
Display:
Display:
Colour Transflective TFT: Display area 111.4
x 83.5mm (4.39 x 3.29in) 320 x 240 pixels.
A-Scan Area 255 x 200 pixels (315 x 200
expanded), 8 colour options and variable
brightness.
Gate
Monitor: Two fully independent
gates for echo monitoring and thickness
measurement. Start and width adjustable
over full range of unit,
amplitude variable from 0 to 100% FSH. Bar
presentation. Positive or negative triggering
for each gate with audible and visual alarms.
Gate
Expansion: Expands range to width
of Gate.
Measurement Modes:
Mode
1: Signal Monitor
Mode
2: Depth and amplitude of first
signal in gate.
Mode
3: Echo-to-Echo distance measurement.
(single gate)
Mode
4: Trigonometric display of beam
path, surface distance and depth of indication,
curve surface correction and X-OFFSET for
probe index. Half skip indication on screen.
Mode
5: Gate to Gate distance measurement.
(independent gates).
Mode
6: T-Min mode for holding minimum
thickness reading. Resolution to 0.01mm
(0.001in) for distance measurement or 1%
FSH for amplitude measurement. Large display
of measurement at top of A-Scan display.
Measurement mode selectable between
peak and flank.
A-Scan
Memory: Maximum of 800 waveforms
can be printed or transferred to a PC using
optional SDMS software.
Panel
Memory: 100 stores for retaining
calibrations.
Thickness
Logging: Storage for 8000 thickness
readings configured either by Block/Location/Number
mode or pre-programmable work sheets in
sequential mode. Readings can be exported
to MS Excel using optional SDMS software.
|
|
AGC:
Automatic Gain Control automatically sets
the signal to a level between 10-90% FSH
with tolerance between 5-20% accuracy.
DAC:
DAC defined by up to 10 points and digitally
drawn on screen. DAC curves meet requirements
of EN 1714, JIS and ASME standards, select
able between -2, -6, -10, -12 and -14dB.
Amplitude read out selectable between %
DAC, % FSH or relative dB.
TCG:
Time Corrected Gain, also known as Swept
Gain. 40dB dynamic range greater than 30dB
per microsecond and up to 10 points may
be used, setting all signals initially to
80% FSH.
Backwall
Echo Attenuation: 0- 40dB attenuation.
AWS:
Built in software for evaluation of defect
indications in accordance with AWS D1.1
structural weld code.
AVG/DGS:
Automatic calculation from transducer data.
Auto-Cal:
Provides automatic calibration from two
echoes.
Clock:
Sets time and date.
Reference
Waveform: This menu displays a
waveform from one of the A-log stores as
a reference or fingerprint display in a
colour different from the active display
highlighting differences from the reference.
Notes:
Alphanumeric labelling for panel and A-log
allows the user to enter Notes for storage
with panel settings and A-scans.
Display
Freeze: For capturing the current
A-scan image.
Active
Peak Memory: Envelope waveform
for echo dynamic pattern determination,
with simultaneous live A-scan display.
Keylock:
Prevents accidental alteration of parameters.
Help
Key: For instant operator guidance
on using the Powerscan 450P.
Language:
Support Supports multiple languages. User
selectable between English, German, Spanish,
French, Dutch, Italian, Russian, Polish,
Czech, Finnish & Hungarian. Others available
on request.
Waveform
Smoothing: Gives a smooth signal
envelope, simulating analogue equipment.
Outputs:
Full bi-directional serial interface to
transfer parameters, thickness readings
and waveform memories. Composite video,
PAL or NTSC compatibility. Analogue proportional
outputs programmable to distance or amplitude
of signal in the gate. Transmitter sync
output.
Front
USB: For connection to printers,
keyboards and PC.
Printers:
Supports any printer with PCL support including
Hp DeskJet, Epson.
Power:
Lithium Ion battery pack 14.4V, 5.0 ampere
hours, gives up to 16 hours duration from
a fully charged pack. Indication of low
battery status. Recharge time 3-4 hrs. Mains
pack option.
Charger:
100 - 240VAC, 50-60Hz.
Transducer
Sockets: BNC or LEMO (factory option)
Environmental:
Case sealed to IP67
Temperature:
Operating -10°C to +55°C (14 to 131°F).
-20°C to +70°C. (-4 to 158°F) survivable.
Storage: -40°C to +75°C. (-40 to = 167°F)
Physical:
Size:
255 x 145 x 145mm (10.0 x 5.7 x 5.7in)
Weight:
2.5kg (5.5lbs) with Li-Ion cells.
Accessories:
Standard
Kit Includes: Powerscan 450P Li-ion
Battery & Battery Charger Fabric Carry Bag
Calibration Certificate Instruction Manual
(EN12668)
Certification:
Calibration
Certificate
|