Search
United Kingdom +44 (0)1908 316345
United States +1 (210) 697 0335
sales@sonatest.com

MASTERSCAN 350M


OverviewSpecifications 
Test Range0 - 1mm (0.05in) up to 0 - 20000 mm (800 in.) at steel
VelocityVariable in 1,2,5 sequence or continuously in 1mm(0.05in) increments. Also from 1 to 5000(μs). Velocity 256 to 16000m/s continuously variable.
Probe Zero0 to 999.999 μs, continuously variable.
Delay Calibrated delay from 0 -20000mm in 0.05 mm steps at steel velocity (0-400in. in 0.002 in. steps).
Gain 0 to 110dB. Adjustable in 0.1, 0.5, 1, 2, 6, 10, 14 and 20dB steps.
Direct access to gain control at all times.
Test ModesPulse echo and transmit/receive.
Pulser100V - 350V (450V MS380) square wave pulser. Pulse width from Spike to 2000ns duration - rise/fall times <5ns into 50 ohms: Width adjustable in 2% of nominal width, minimum 1ns maximum 40ns.
ActiveEdge TMUnique active pulse control for enhanced near surface resolution and signal response. Replaces traditional damping control.
P.R.FSelectable 5 to 5000 Hz.
Update Rate60Hz (NTSC Mode); 50Hz (PAL Mode).
RectificationFull wave, positive or negative half wave and unrectified rf.
Frequency Range6 narrow bands centred at 0.5 MHz, 1MHz, 2.25MHz, 5MHz, 10MHz and 15MHz. Broad band at 2 MHz to 22MHz (-6dB) and 1MHz to 35 MHz (-20dB).
System LinearityVertical = 1% Full Screen Height (FSH). Amplifier Accuracy ± 0.1dB. Horizontal ±0.4% Full Screen Width (FSW).
Reject80% linear reject. LED warning light when selected.
UnitsMetric (mm), inch (in) or time (μs).
DisplayColour Transflective TFT: Display area 111.4 x 83.5 mm (4.39 x 3.29 in) 320 x 240 pixels. A-Scan Area 255 x 200 pixels (315 x 200 expanded), 8 colour options and variable brightness.
GateMonitor Two fully independent gates for echo monitoring and thickness measurement. Start and width adjustable over full range of unit, amplitude variable from 0 to 100% FSH. Bar presentation. Positive or negative triggering for each gate with audible and visual alarms.
Gate ExpansionExpands range to width of Gate 1.
Gate MonitorDelay Selectable 0.6 seconds delay on gate 2 negative monitor
tracking.
Measurement Modes 
Mode 1Signal Monitor
Mode 2Depth and amplitude of first signal in gate.
Mode 3Echo-to-Echo distance measurement. (single gate)
Mode 4Trigonometric display of beam path, surface distance and depth of indication, curve surface correction and X-OFFSET for probe index. Half skip indication on screen.
Mode 5Gate to Gate distance measurement. (independent gates).
Mode 6T-Min mode for holding minimum thickness reading. Resolution to 0.01mm (0.001in) for distance measurement or 1% FSH for amplitude measurement. Large display of measurement at top of A-Scan display. Measurement mode selectable between peak and flank.
A-Scan MemoryMaximum of 800 waveforms can be printed or transferred to a PC using optional SDMS software.
Panel Memory100 stores for retaining calibrations.
Thickness LoggingStorage for 8000 thickness readings configured either by Block/Location/Number mode or pre-programmable work sheets in sequential mode. Readings can be exported to MS Excel using optional SDMS software.
AGCAutomatic Gain Control automatically sets the signal to a level between 10-90% FSH with tolerance between 5-20% accuracy.
DACDAC defined by up to 10 points and digitally drawn on screen. DAC curves meet requirements of EN 1714, JIS and ASME standards, select able between -2, -6, -10, -12 and -14dB. Amplitude read out selectable between % DAC, % FSH or relative dB.
TCGTime Corrected Gain, also known as Swept Gain. 40dB dynamic range greater than 30dB per microsecond and up to 10 points may be used, setting all signals initially to 80% FSH.
Back wall Echo Attenuation0- 40dB attenuation.
AWSBuilt in software for evaluation of defect indications in accordance with AWS D1.1 structural weld code.
AVG/DGSAutomatic calculation from transducer data.
APIFlaw sizing complying with API 5UE
Auto-CalProvides automatic calibration from two echoes.
ClockSets time and date.
Reference WaveformThis menu displays a waveform from one of the A-log stores as a reference or fingerprint display in a colour different from the active display highlighting differences from the reference.
NotesAlphanumeric labelling for panel and A-log allows the user to enter Notes for storage with panel settings and A-scans.
Display FreezeFor capturing the current A-scan image.
Active Peak MemoryEnvelope waveform for echo dynamic pattern determination, with simultaneous live A-scan display.
KeylockPrevents accidental alteration of parameters.
Help KeyFor instant operator guidance on using the Masterscan Series.
Language Support Supports multiple languages. User selectable between English, German, Spanish, French, Dutch, Italian, Russian, Polish, Czech, Finnish & Hungarian. Others available on request.
Waveform SmoothingGives a smooth signal envelope, simulating analogue equipment.
OutputsFull bi-directional serial interface to transfer parameters, thickness readings and waveform memories. Composite video, PAL or NTSC compatibility. Analogue proportional outputs programmable to distance or amplitude of signal in the gate. Transmitter sync output
Front USBFor connection to printers, keyboards and PC.
Printers Supports Hp DeskJet, Epson.
PowerLithium Ion battery pack 14.4V, 5.0 ampere hours, gives up to 16 hours duration from a fully charged pack. Indication of low battery status. Recharge time 3-4 hrs. Mains pack option.
Charger100 - 240 VAC, 50-60 Hz.
Transducer SocketsBNC or LEMO (factory option)
EnvironmentalCase sealed to IP67
TemperatureOperating -10°C to +55°C (14 to 131°F).
-20°C to +70°C. (-4 to 158°F) survivable.
Storage:-40° to +75°C. (-40 to = 167°F)
Size255 x 145 x 145mm (10.0 x 5.7 x 5.7 in)
Weight2.5kg (5.5lbs) with Li-Ion cells.
Standard KitLi-ion Battery & Battery Charger
Fabric Carry Bag
Calibration Certificate
Instruction Manual (EN12668)
 
Get a competitive quotation
Get a competitive quotation
(price) on this product from
Sonatest or one of our many
authorized distributors.


Request more information about this product. This may be technical queries, applications that this product is used in or any other information you require related to this product.
Request more information about
this product. This may be